Chinese Journal of Endourology(Electronic Edition) ›› 2025, Vol. 19 ›› Issue (06): 778-783. doi: 10.3877/cma.j.issn.1674-3253.2025.06.015

• Experiment Research • Previous Articles    

Study on the effectiveness of a novel dual channel laparoscopic smoke evacuation device

Chao Zhi1, Bin Cao2, Liang Xiao3, Jia Wu3, Wenjun Xue3, Chao Zhang1, Fei Guo1,(), Bo Yang1   

  1. 1Department of Urology, the First Affiliated Hospital of Naval Medical University, Shanghai 200433, China
    2Department of Urology, Jingjiang People's Hospital, Jiangsu 214504, China
    3Shanghai Oriental Institute of Medical Innovation, Shanghai 200433, China
  • Received:2025-04-07 Online:2025-12-01 Published:2025-11-24
  • Contact: Fei Guo

Abstract:

Objective

A laparoscopic novel dual-channel smoke evacuation device was designed, and its smoke exhaust effect was verified.

Methods

In the in vitro transparent laparoscopic simulation box, the chicken fat was grasped by ultrasonic knife and bipolar electrocoagulation forceps, and the novel dual-channel smoke evacuation device installed group (study group) and the traditional smoke extraction device installed group (control group) were compared. The smoke particle count in the simulated box was measured. The laparoscopic lens was used to observe the experimental operation at different distance, and the number of times the lens contaminated during 100 operations was recorded.

Results

In the smoke evacuated experiment, the number of particles measured in the study group about 0.3 μm particles [(1 329 247 ± 159 800) vs (24 123 431 ± 9 366 698), P=0.006], 0.5 μm particles [(339 856 ± 96 857) vs (3 037 262 ± 1 038 992), P<0.001], and 1.0 μm particles [(60 542 ± 38 500) vs (779 007 ± 96 318), P<0.001], 2.5 μm particles [(4 938 ± 1 938)vs (89 846 ± 20 408), P<0.001], 5.0 μm particles [(1 365 ± 414) vs (42 649 ± 8 489), P<0.001], 10.0 μm particles [(36 ± 17) vs (2 797 ± 1 094), P<0.001] were significantly lower than those in the control group. In the experiment of lens anti-pollution effect, the number of contaminated operations observed by the study group at middle distance (5 cm) (13/100 vs 27/100 times, P=0.013) and the number of contaminated operations observed by the lens at close distance (3 cm) (18/100 vs 40/100 times, P<0.001) were significantly fewer than those in the control group.

Conclusion

The novel dual-channel smoke evacuation device can effectively evacuate smoke and maintain a clear surgical field of view during laparoscopic surgery.

Key words: Laparoscopy, Smoke extraction device, Surgical smoke, Dual-channel, Equipment design, Occupational exposure